ASTM D402
This procedure measures the amount of the more volatile constituents in cutback asphaltic products. The properties of the residue after distillation are not necessarily characteristic of the bitumen used in the original mixture nor of the residue, which may be left at any particular time after field application of the cutback asphaltic product. The presence of silicone in the cutback may affect the distillation residue by retarding the loss of volatile material after the residue has been poured into the residue container.
LT/CB-106000/M:
• Electric heater supported on a height-adjustable platform, with main switch and heater power regulator.
• Chimney 117 mm diameter with an insulated metal shield and insulated cover split into two halves.
• Distillation Flask, 500 ml, side-arm.
• Adapter made in glass with an angle of 105°.
• Water jacket condenser and nozzle extensor 450 mm.
• Cylinder 100 ml capacity.
• Stand with a clamp to hold the glass cooler.
ASTM D86 - Distillation of Petroleum Products at Atmospheric Pressure: This test method covers the atmospheric distillation of petroleum products using a laboratory batch distillation unit to determine the boiling range characteristics of such products as natural gasoline, light and middle distillates, automotive spark-ignition engine fuels, aviation types of gasoline, aviation turbine fuels, 1-D and 2-D regular and low sulfur diesel fuels, unique petroleum spirits, naphthas, white spirits, kerosines, and Grades 1 and 2 burner fuels quantitatively. The test method is designed to analyze distillate fuels; it does not apply to products containing appreciable quantities of residual material.
ASTM D216 (obs.), ASTM D447 (obs.) - Distillation Test Method.
ASTM D447 (obs.) - Test Method for Distillation of Plant Spray Oils.
ASTM D850—Distillation of Industrial Aromatic Hydrocarbons and Related Materials: This test method covers industrial aromatic hydrocarbons and related materials with relatively narrow boiling ranges from 30 to 250°C.
ASTM D1078, IP 195—Distillation Range of Volatile Organic Liquids: This test method covers the determination of the distillation range of liquids boiling between 30 and 350°C that are chemically stable during the distillation process, using manual or automatic distillation procedures. It applies to organic liquids such as hydrocarbons, oxygenated compounds, chemical intermediates, and blends.
ASTM E 133, IP 123, DIN 51751, ISO 3405 - Standard Specification for Distillation Equipment: This specification covers distillation equipment used in the following ASTM test methods: D86, D216, D447, D850, and D1078.
LT/HCU-99000/M:
The manual instrument for distillation is composed of the following:
• Structure entirely made in stainless steel
• Front panel including manual controls for heating power, fan activation, and main power supply
• Plate supported by a base whose height is adjustable with an elevation mechanism controlled by an external knob
• Electric heater 1200 Watt with ceramic-glass plate support
• Wide toughed glass squared window and stainless steel cover with a hole for flask neck
• Cooling fan manually activated for cooling down the glass after analysis
• Condensing unit entirely made in stainless steel with double chamber insulation:
- Condensing tube made of stainless steel
- Insulated cover with handle and hole for the accommodation of the thermometer with relevant support and liquid level indicator
- Rear connection for coolants circulation, overflow tube, and atmospheric drain cock
- White background panel for easier reading of the receiver glass cylinder level mark
LT/HCU-99000/M+:
The manual instrument for distillation is composed of the following:
• Structure entirely made in stainless steel
• Front panel including manual controls for heating power, fan activation, and main power supply
• Plate supported by a base whose height is adjustable with an elevation mechanism controlled by an external knob
• Controlled by a digital thermoregulator with PT100 A class temperature sensor
• PID range from ambient to +450°C, resolution 0,1°C
• Electric heater 1200 Watt with ceramic-glass plate support
• Wide toughed glass squared window and stainless steel cover with a hole for flask neck
• Cooling fan manually activated for cooling down the glass after analysis
• Condensing unit entirely made in stainless steel with double chamber insulation:
- Condensing tube made of stainless steel
- Insulated cover with handle and hole for the accommodation of the thermometer with relevant support and liquid level indicator
- Rear connection for coolants circulation, overflow tube, and atmospheric drain cock
- White background panel for easier reading of the receiver glass cylinder level mark
LT/RDS-900/SA: Semi-automatic instrument for distillation is composed by:
• Structure entirely made in stainless steel
• Front panel including manual controls for heating power, fan activation, and main power supply
• Plate supported by a base whose height is adjustable with an elevation mechanism controlled by an external knob
• Controlled by a digital thermoregulator with PT100 A class temperature sensor
• PID range from ambient to +450°C, resolution 0,1°C
• three programmable set points for Gasoline, Kerosene, and Gasoil for semi-automatic operation mode
• Infrared heaters 1300 Watt with ceramic-glass plate support
• Fire extinguisher system composed by:
- solenoid valve
- red emergency push button
- dedicated line internally placed with holes for the emission of the fire extinguisher product with a joint for the external connection
ASTM D244: These test methods and practices cover the examination of asphalt emulsions composed principally of a semisolid or liquid asphaltic base, water, and an emulsifying agent.
ASTM D6997: This test method quantitatively determines residue and oil distillate in emulsified asphalts composed of a semisolid or liquid asphaltic base, water, and an emulsifying agent.
LT/RD-271000/M:
Residue by Distillation Apparatus for Emulsified Asphalts ASTM D244 - D6997:
• Aluminium alloy boiler with an annular gas lamp for heating
• Connection glass tube with protection shield
• Glass condenser for water circulation
• Graduated cylinder 100 ml
• Thermometer ASTM 7C
• Supporting ring
• Bases with rods
• Pliers